Shopping security
In summary, the Rohde & Schwarz RTH1024 is a 4-channel CAT IV 600 V / CAT III 1000 V rated handheld digital oscilloscope built on the R&S Scope Rider RTH platform. 4 oscilloscope channels, with 200 MHz analog bandwidth, 5 GSa/s peak real-time sampling, a 10-bit ADC (up to 16-bit with high-resolution decimation), and calculated rise time of < 1.75 ns. The R&S RTH-K15 history and segmented memory option ships with the base unit, and the platform is software-upgradeable to higher bandwidth tiers via keycode without returning the instrument.
The RTH1024 is used for waveform debug and measurement on energized industrial and electrical equipment — motor drives, switching power supplies, photovoltaic inverters, distribution panels, automotive electronics, and field service of installed systems. Its fully isolated inputs and CAT IV 600 V / CAT III 1000 V rating allow direct connection to live AC mains and three-phase distribution circuits without an isolation transformer.
Oscilloscope bandwidth governs the highest signal frequency that can be measured with rated accuracy. The standard rule is to choose an instrument with 3 to 5 times the bandwidth of the highest signal frequency in the device under test. At 200 MHz, the RTH1024 delivers a calculated rise time of < 1.75 ns, suitable for the digital edges, switching transients, and PWM waveforms encountered in modern industrial and power-electronics environments.
Quantitative anchors from the datasheet: peak real-time sampling of 5 GSa/s on a single active channel, DC gain accuracy ±1.5% of reading at input sensitivities above 5 mV/div, channel-to-channel isolation greater than 40 dB (measured) at matched input sensitivity, common-mode rejection greater than 100 dB at DC and AC ≤ 100 kHz (measured), timebase accuracy ±10 ppm, and waveform acquisition rate up to 50,000 waveforms/s. Input impedance is 1 MΩ ± 1% in parallel with 12 pF ± 2 pF (measured).
The RTH1024 is built and certified to IEC/EN/DIN EN 61010-1, IEC/EN/DIN EN 61010-2-030, UL/CSA 61010-1, UL/CSA 61010-2-030. EMC compliance follows CISPR 11 / EN 55011 Group 1 Class A and IEC/EN 61326-1. Environmental qualification covers MIL-PRF-28800F class 3 shock and vibration and IP51 ingress protection per IEC 60529. The instrument carries VDE, CCSAUS, and KC certifications.
| Item | Description |
|---|---|
| Voltage probes | One 500 MHz, 10:1, 600 V CAT IV passive voltage probe per oscilloscope input channel |
| microSD card | Installed in the instrument; user data and settings storage |
| Battery pack | Lithium-ion smart battery, 72 Wh, 11.25 V |
| Power supply | External adapter with plugs for EU, CH, UK, US, CAN, China, Australia |
| RTH-K15 option | History and segmented memory option installed with the base unit |
| Compact manual | Printed quick-start reference |
| Input channels | 4 oscilloscope channels; all inputs floating and fully isolated |
| Input impedance | 1 MΩ ± 1% || 12 pF ± 2 pF (meas.) |
| Analog bandwidth (−3 dB) | ≥ 200 MHz |
| Lower frequency limit (AC) | < 8 Hz (meas.) |
| Rise time (calculated) | < 1.75 ns |
| ADC resolution | 10 bit; up to 16 bit with high-resolution decimation |
| DC gain accuracy | ±1.5% of reading (>5 mV/div); ±2% (>2 mV/div to 5 mV/div); ±2.5% (2 mV/div) |
| Input sensitivity | 2 mV/div to 100 V/div in 1, 2, 4, 5 steps per decade |
| Max input voltage (BNC) | CAT IV 300 V (RMS), 424 V (peak); derates 20 dB/decade above 500 kHz |
| CMRR | > 100 dB at DC and AC ≤ 100 kHz (meas.) |
| Timebase range | 1 ns/div to 500 s/div |
| Timebase accuracy | ±10 ppm |
| Max real-time sampling | 5 GSa/s (1 active channel); 2.5 GSa/s (2 channels); 1.25 GSa/s (3 or 4 channels) |
| Max acquisition length | 500 ksample (1 ch, sample mode); 250 ksample (2 ch, sample mode); 125 ksample (3 or 4 ch, sample mode) |
| Acquisition modes | Sample, high resolution, peak detect, envelope, average (2 to 8192 waveforms) |
| Waveform acquisition rate | Max 50,000 waveforms/s |
| Trigger sources | CH1, CH2, CH3, CH4 |
| Trigger modes | Auto, normal, single |
| Base trigger types | Edge, glitch, width, TV/video (PAL/NTSC/SECAM/PAL-M) |
| With RTH-K19 | Pattern, state, runt, slew rate, window, data2clock, serial pattern, timeout, interval, SDTV/HDTV |
| Hold-off | 8 ns to 10 s; 1 to 1,000,000,000 events |
| Display | 7.0" LC TFT, 800 × 480 (WVGA) |
| Interfaces | USB host (Type A, 2.0); USB device (mini-B); LAN 10/100BASE-T (RJ-45); microSD slot |
| Battery / operating time | Lithium-ion smart battery, 72 Wh, 11.25 V; approx. 4 h operating; approx. 4 h charging (off) |
| Operating temperature | 0 °C to +50 °C (battery); 0 °C to +40 °C (adapter) |
| IP rating | IP51 per IEC 60529 |
| Shock / vibration | 40 g per MIL-STD-810E method 516.4; MIL-PRF-28800F class 3 |
| Dimensions (W × H × D) | 201 × 293 × 74 mm (7.91 × 11.54 × 2.91 in) |
| Weight (with battery) | 2.4 kg (5.3 lb) nom. |
| Calibration interval | 1 year |
Important note on probes. Maximum input voltage at the BNC inputs derates above 500 kHz per the published input rating curve. To preserve full CAT IV 600 V / CAT III 1000 V isolation to the probe tip, use R&S RT-ZI10 or RT-ZI11 isolated probes. Standard 600 V CAT IV passive probes ship with the base unit.
Note on calibration. Manufacturer-recommended calibration interval is 1 year. Optional accredited calibration coverage is available via R&S extended-warranty programs (CW1/CW2, AW1/AW2) — contact your Test Architect for details.
This webpage was written by a human with an A.I. "Intern", which may contain errors. Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.